Analog Discovery Pro (ADP5250) Specifications
These specifications are valid following 30 minutes of warmup and are typical at 25 °C unless otherwise noted.
Mixed Signal Oscilloscope
Analog Input Channels
Supports the Oscilloscope, Data Logger, Spectrum Analyzer, Network Analyzer, Impedance Analyzer, and Script Editor instruments.
Vertical System
Number of Channels | 2 single ended, non-isolated | |
---|---|---|
Bandwidth (-3 dB) | 100MHz | |
Resolution | 8 bits | |
Accuracy | ±2% of input ±1% full scale (V peak to peak) | |
Input Coupling | AC, DC | |
Vertical Sensitivity (range) | 10 mV / div (100 mV peak to peak) 20 mV / div (200 mV peak to peak) 50 mV / div (400 mV peak to peak) 100 mV / div (1 V peak to peak) 200 mV / div (2 V peak to peak) 500 mV / div (4 V peak to peak) 1 V / div (10 V peak to peak) 2 V / div (20 V peak to peak) 5 V / div (40 V peak to peak) |
|
Input Impedance | 1 MΩ ││ 20 pF | |
Acquisition Modes | Sample, peak detect, averaging | |
直流偏置Range | 10 mV/div, 20 mV/div, 50 mV/div ranges: ±5 V 100 mV/div, 200 mV/div, 50 mV/div, 1 V/div, 2 V/div, 5 V/div ranges: ±20 V |
Horizontal System
Maximum Sampling Rate | 1GS/ssingle channel 500MS/sper channel, dual channel |
|
---|---|---|
Maximum Record Length | 1 MS per channel |
Digital Channels / Logic Analyzer
Vertical System
Number of Channels | 341) | |
---|---|---|
Maximum Input Frequency | 100MHz | |
Input Voltage | 0 V to 5 V | |
Input Current | ≤50 µA | |
Input Threshold | Programmable, 0 V to 2.0 V | |
Threshold Accuracy | 350 mV | |
Input Impedance | 100 kΩ ││ 7.5 pF pulled to -2.0 V to +6.5 V, varies with input threshold setting | |
Additional / Internal Channels | Digital I/O lines, function generator start, external trigger (TRIG), power line frequency |
Note:Mixed signal oscilloscope digital channels are designed to withstand accidental overvoltage from signals on the ADP5250. They are not recommended for use with signals likely to exceed 0 V to 5 V in normal operation.
Horizontal System
Timing Mode Sample Rate | 1GS/s(down to ~15kS/s) | |
---|---|---|
Maximum External Sample Clock Rate | 100MHz | |
Record Length (Typical) | 1 MS | |
Record Length (Minimum) | 4 kS | |
Decimation | External Sample Clock, 1:1, 2:1, and n*4:1 where n is an integer | |
Maximum Sample Compression | 215to 1 |
Triggering
Trigger Modes | Normal, Auto, Force, Single | |
---|---|---|
Trigger Sources | Oscilloscope analog channels Oscilloscope digital channels Function generator start Digital I/O lines External trigger (TRIG) Power line frequency |
|
Trigger Types | Analog | Edge with hysteresis |
Digital | Edge, pattern, glitch, level | |
Trigger Resolution | Analog / Oscilloscope | 667 ps |
Digital / Logic Analyzer | 1纳秒 | |
Trigger Export | Available through external trigger (TRIG) |
Waveform Generator (WaveGen)
Supports the Waveform Generator, Network Analyzer, Impedance Analyzer, and Script Editor instruments.
Waveforms | Sine, square, ramp / triangle, DC, arbitrary | |
---|---|---|
Update Rate | 125MS/s | |
Resolution | 14 bits | |
Number of Channels | 1 | |
Output Impedance | 50 Ω | |
Switchable Filter | 36MHz低通滤波器、7-pole椭圆 | |
Sine | Maximum Frequency | 20MHz |
Total Harmonic Distortion (1MHz) | -55 dBc | |
Total Harmonic Distortion (10MHz) | -50 dBc | |
Spurious Free Dynamic Range (SFDR) | -70 dB at 1MHz(non-harmonic) | |
Phase Noise (1MHz) | -115 dBc/Hzat 10kHzoffset | |
Square | Maximum Frequency | 5MHz |
Rise / Fall Time | <20 ns (10% to 90%) | |
Overshoot | <5% | |
Jitter | 8 ns cycle-to-cycle | |
Ramp / Triangle Maximum Frequency | 1MHz | |
Accuracy (with >10 kΩ load) | Amplitude (1kHzsine) | ±(1% of output value ±5 mV) |
DC | ±(1% of output value ±7.5 mV) | |
Output Range | 50 Ω | ±6 V |
Hi-Z (>10 kΩ) | ±12 V | |
直流偏置 | 50 Ω | ±6 V |
Hi-Z (>10 kΩ) | ±12 V | |
Frequency | Accuracy | ≤100 ppm |
Resolution | 1 µHz | |
Arbitrary Waveforms | Points | 1 MS |
Sample Rate | 125MS/s | |
Protection | Short-circuit protected |
Note:The combination of signal amplitude and DC offset cannot exceed the output range specifications. The impedances listed are the loads applied by the user to the Waveform Generator output.
Digital I/O
Supports the Static I/O and Script Editor instruments.
Number of Channels | 8 | |
---|---|---|
Direction Control | Input or output, software-selectable | |
Logic Level | 5 V compatible TTL input 3.3 V LVTTL output |
|
Drive Strength | 4 mA | |
Input Voltage | 0 to 5 V | |
DIO Channel Pull Resistors | 10 kΩ, pull-down on dig/<0..7> 10 kΩ, configurable pull-up to 3.3 V on dig/<6,7> |
Note:Digital I/O lines are designed to withstand accidental overvoltage from signals on the ADP5250 or similar devices. They are not recommended for use with signals likely to exceed 0 V to 5 V in normal operation.
Digital Power Supply
Voltage | 3.3 V ± 10% |
---|---|
Current | 20 mA |
Digital Multimeter
Functions | DC Voltage, AC Voltage, DC current, AC current, resistance, diode, continuity | |
---|---|---|
Resolution | 5½ digits | |
Sample Rate | 5S/s |
Caution:Do not use this device for connection to signals or for measurements within Measurement Categories III or IV. For more information about Measurement Categories, refer to theSafety Voltagessection.
Input Protection
Resistance, Diode | Up to 300 V DC | |
---|---|---|
DC and AC Voltage | Up to 300 V DC or 265 V AC (rms) 400 V AC (peak) | |
DC and AC Current | DMM A Current Connector Fuse | Internal ceramic fuse, 10 A 250 V, time-delay, 5 x 20 mm, T 10A H 250V (Bussman part number D505H-10-R at www.cooperindustries.com) |
DMM mA Current Connector Fuse | Internal ceramic fuse, 1.25 A 250 V, time-delay, 5 x 20 mm, T 1.25A H 250V (Bussman part number S505H-1.25-R at www.cooperindustries.com) |
Caution:Fuses are located on bottom of device underneath door. Use Phillips #1 screwdriver for removal. Ensure all hazardous voltages are disconnected from the device prior to removal of door.
Fuse:When the fuse symbol is marked on a device, take proper precautions.
Maximum common-mode voltage | 300 V DC or AC (rms) |
---|
DC
DC Voltage Accuracy
Range | Input Impedance | 1-Year Accuracy ±(% of Reading + % of Range) | Temperature Coefficient ±(% of Reading + % of Range)/C° | |
---|---|---|---|---|
100 mV* | <10 GΩ, 10 MΩ | 0.015 + 0.005 | 0.001 + 0.0005 | |
1 V | >10 GΩ, 10 MΩ | 0.015 + 0.005 | 0.001 + 0.0005 | |
10 V | >10 GΩ, 10 MΩ | 0.015 +0.005 | 0.001 + 0.0005 | |
100 V | 10 MΩ | 0.035 + 0.005 | 0.001 + 0.0005 | |
300 V | 10 MΩ | 0.035 + 0.005 | 0.001 + 0.0005 |
* Add 15 µV if not immediately following offset null.
DC Current Accuracy
Range | Burden Voltage | 1-Year Accuracy ±(% of Reading + % of Range) | Temperature Coefficient ±(% of Reading + % of Range)/C° | |
---|---|---|---|---|
10 mA | <0.03 V | 0.070 + 0020 | 0.0035 + 0.0010 | |
100 mA | <0.3 V | 0.070 + 0.003 | 0.0020 + 0.0010 | |
1 A | <0.03 V | 0.130 +0.025 | 0.0065 + 0.0010 | |
10 A* | <0.3 V | 0.130 + 0.004 | 0.0045 + 0.0010 |
* 30 seconds on, 30 seconds off. Add 300 ppm for currents >2.2 A. After measuring >5 A, Wait two minutes to get full accuracy in the 1 A range.
DC Resistance Accuracy (2-Wire)*, 1V Open Circuit Voltage
Range | Short-Circuit Current | 1-Year Accuracy ±(% of Reading + % of Range) | Temperature Coefficient ±(% of Reading + % of Range)/C° | |
---|---|---|---|---|
100 Ω | 170 µA | 0.018 + 0.050 | 0.0010 + 0.0005 | |
1 kΩ | 170 µA | 0.018 + 0.005 | 0.0010 + 0.0005 | |
10 kΩ | 70 µA | 0.018 + 0.005 | 0.0010 + 0.0005 | |
100 kΩ | 10 µA | 0.018 + 0.005 | 0.0010 + 0.0005 | |
1 MΩ | 1.1 µA | 0.035 + 0.005 | 0.0040 + 0.0005 | |
10 MΩ | 1.1 µA | 0.150 + 0.005 | 0.0100 + 0.0005 | |
100 MΩ | 1.1 µA | 1.3 + 0.005 | 0.1000 + 0.0005 |
* Perform offset nulling
Caution:The input terminals of the DMM are not protected for electromagnetic interference. As a result, the DMM may experience reduced measurement accuracy or other temporary performance degradation when connected to unshielded test leads in an environment with radiated or conducted radio frequency electromagnetic interference.
DC Continuity Accuracy Range | 100 Ω | |
---|---|---|
DC Diode Test Range | 2 V | |
Effective Common-Mode Rejection-Mode Ratio (CMRR), 1 kΩ resistance in LO lead | >100 dB | |
Normal-Mode Rejection Ratio (NMRR), 50/60Hz± 0.1% | >100 dB | |
Overrange | 105% of range except 300 V |
AC
AC Voltage Accuracy
Range (rms) | Peak Voltage | Frequency | 1-Year Accuracy ±(% of Reading + % of Range) | Temperature Coefficient ±(% of Reading + % of Range)/C° | |
---|---|---|---|---|---|
100 mV, 1 V, 10 V, 100 V, 265 V | ±210 mV, ±2.1 V, ±21 V, ±210 V, ±400 V | 20Hzto 45Hz | 0.91 + 0.10 | 0.01 + 0.005 | |
45Hzto 65Hz | 0.30 + 0.05 | 0.01 + 0.005 | |||
65Hzto 1kHz | 0.21 + 0.05 | 0.01 + 0.005 | |||
1kHzto 5kHz | 0.12 + 0.05 | 0.01 + 0.005 | |||
5kHzto 20kHz | 0.35 + 0.05 | 0.01 + 0.005 |
AC Current Accuracy
Range | Short-Circuit Current | Burden Voltage (rms) | Frequency | 1-Year Accuracy ±(% of Reading + % of Range) | Temperature Coefficient ±(% of Reading + % of Range)/C° | |
---|---|---|---|---|---|---|
5 mA | ±10.5 mA | <0.02 | 20Hzto 1kHz | 0.20 + 0.01 | 0.01 + 0.005 | |
5 mA | ±10.5 mA | <0.02 | 1kHzto 5kHz | 0.60 + 0.01 | 0.01 + 0.005 | |
50 mA | ±105 mA | <0.2 V | 20Hzto 1kHz | 0.20 + 0.01 | 0.01 + 0.005 | |
50 mA | ±105 mA | <0.2 V | 1kHzto 5kHz | 0.50 + 0.01 | 0.01 + 0.005 | |
500 mA | ±1.05 A | <0.02 V | 20Hzto 1kHz | 0.15 + 0.01 | 0.01 + 0.005 | |
500 mA | ±1.05 A | <0.02 V | 1kHzto 5kHz | 0.50 + 0.01 | 0.01 + 0.005 | |
5 A | ±1.05 A | <0.02 V | 20Hzto 1kHz | 0.25 + 0.03 | 0.01 + 0.005 | |
5 A | ±1.05 A | <0.02 V | 1kHzto 5kHz | 0.60 + 0.03 | 0.01 + 0.005 |
DC Power Supply
Outputs | 0 V to +6 V | 0 A to 1 A | |
---|---|---|---|
0 V to +25 V | 0 mA to 500 mA (isolated) | ||
0 V to -25 V | 0 mA to 500 mA (isolated) |
Note:the +25 V and –25 V channels are bank isolated from ground but not from each other.
DC Accuracy / Resolution
Output | Type | +6 V | +25 V | -25 V | |
---|---|---|---|---|---|
DC Output | Voltage | 0 V to +6 V | 0 V to +25 V | 0 V to -25 V | |
Current | 1 A | 500 mA | 500 mA | ||
Programming Accuracy, ±(% of reading + offset) | Voltage | 0.1% + 5 mV | 0.15% + 20 mV | 0.15% + 20 mV | |
Current | 0.2% + 10 mA | 0.15% + 4 mA | 0.15 + 4 mA | ||
Readback Accuracy, ±(% of reading + offset) | Voltage | 0.1% + 5 mV | 0.15% + 20 mV | 0.15% + 20 mV | |
Current | 0.2% + 10 mA | 0.15% + 4 mA | 0.15% + 4 mA | ||
Programming Resolution | Voltage | 1.7 mV | 6.5 mV | 6.5 mV | |
Current | 0.30 mA | 0.15 mA | 0.15 mA | ||
Readback Resolution | Voltage | 0.41 mV | 1.7 mV | 1.7 mV | |
Current | 70 µA | 35 µA | 35 µA | ||
Load Regulation ±(% of reading + offset) | Voltage | 0.01% + 25 mV | 0.03% + 5 mV | 0.03% + 5 mV |
Overvoltage Protection | 30 V (25 V channels) and 10 V (6 V channel) | |
---|---|---|
Reverse Voltage Protection | Reverse clamp diode, protected by self-resetting fuse |
External Trigger (TRIG)
Direction Control | Input or output, software-selectable | |
---|---|---|
Logic Level | 5 V compatible TTL input, 3.3 V LVLTTL output | |
Drive Strength | 4 mA | |
Input Voltage | 0 V to 5 V |
Note:The external trigger line is designed to withstand accidental overvoltage from signals on the ADP5250 or similar devices. It is not recommended for use with signals likely to exceed 0 V to 5 V in normal operation.
Features
Spectrum Analyzer
Frequency Range | DC to maximum bandwidth of scope | |
---|---|---|
Display Modes | Magnitude, average, peak hold, min hold, count | |
Y Axis | Logarithmic (dbV, dBu, dBm) or linear (volts) | |
X Axis | Linear or logarithmic | |
Windowing Functions | Rectangular, Triangular, Hamming, Hann, Cosine, Blackman-Harris, Flat Top, Kaiser |
Network Analyzer
Frequency Range | 1兆赫兹10MHz, up to 10k samples per decade | |
---|---|---|
Display Modes | Magnitude, Phase | |
Y Axis | Linear or logarithmic | |
X Axis | Linear or logarithmic | |
Plots | Time, FFT, Nichols, Nyquist |
Math Channels
Operations | addition “+”, subtraction “-”, multiplication “*”, division “/”, reminder “%”. | |
---|---|---|
Brackets | parentheses (), square brackets [] | |
Constants | Exp, Ln, Log, Pi | |
Functions | Logarithm, power, minimum, maximum, square root, sine, cos, tan, arccos, arctan, arctan2, absolute value, round, floor, ceil | |
Operands | All analog and digital input channels, reference waveforms, time, constants, Pi |
Automatic Measurements (Analog Channels Only)
Oscilloscope Mode | DC to maximum bandwidth of scope | |
---|---|---|
Spectrum Mode | Magnitude, average, peak hold, min hold, count | |
Statistics | Logarithmic (dbV, dBu, dBm,) or linear (volts) |
Connectivity
Wired USB Interface
USB Specification | USB 2.0 Hi-Speed |
---|
Power Requirements
Voltage Input Range | 100 V AC to 240 V AC, ±10%, 50/60Hz | |
---|---|---|
Power Consumption | 100 W maximum | |
电源输入连接器 | IEC C13 power connector | |
Power Disconnect | The AC power cable provides main power disconnect. Do not position the equipment so that it is difficult to disconnect the power cable. Depressing the front panel power button does not inhibit the internal power supply. |
Physical Characteristics
Enclosure | 25.40 cm x 19.05 cm x 7.77 cm (10.00 in x 7.50 in x 3.06 in) |
|
---|---|---|
Weight | 2.05 kg (4 lb 8.3 oz) | |
Security Cable Slot | 1, complies with Kensington security slot dimensions |
Note:If the device needs to be cleaned, wipe with a dry towel.
Note:Use the ADP5250 in a horizontal orientation. Allow at least 10.16 cm (4.0 in.) of clearance in front and behind the ADP5250 for USB, power, and common connector cabling.
Safety Voltages
Connect only voltages that are within these limits.
地球DMM隔离电压(地面)
Channel-to-earth Ground
Continuous | 300 V, Measurement Category II | |
---|---|---|
Withstand | 3000 V (RMS), verified by a 5 s dielectric withstand test |
Measurement Category II is for measurements performed on circuits directly connected to the electrical distribution system. This category refers to local-level electrical distribution, such as that provided by a standard wall outlet, for example, 115 V for U.S. or 230 V for Europe.
Caution:Do not connect the ADP5250 hardware to signals or use for measurements within Measurement Categories III or IV. |
DC Power Supply Isolation Voltages
+25V and -25V to earth ground, continuous | 60 VDC, Measurement Category 1 |
---|
Note:测量类别猫我和猫O(其他)equivalent. The input circuits are not intended for direct connection to the MAINs building installations of Categories CAT II, CAT III, or CAT IV.
Environmental
Operating Temperature | 0 °C to 40 °C | |
---|---|---|
Storage Temperature | -20 °C to 60 °C | |
Operating Humidity | 10% to 90% RH, non-condensing DMM full accuracy at 10% to 80% |
|
Storage Humidity | 5% to 95% RH, non-condensing | |
Cooling | Forced air circulation (positive pressurization) through a fan. Fan speed automatically adjusts according to operating conditions. Intake and exhaust locations are on rear of device. Ensure that the intake and exhaust locations are not obstructed. | |
Pollution Degree | 2 | |
Maximum Altitude | 2,000 m |
Note:For indoor use only.
Shock and Vibration
操作冲击 | 30 g peak, half sine, 11 ns pulse (Tested in accordance with IEC 60068-2-27. Test profile developed in accordance with MIL-PRF-28800F.) | |
---|---|---|
Random Vibration (Operating) | 5Hzto 500Hz, 0.3 grms | |
Random Vibration (Nonoperating) | 5Hzto 500Hz, 2.4 grms(Tested in accordance with IEC 60068-64. Nonoperating test profile exceeds the requirements of MIL-PRF-28800F, Class 3.) |
Safety
This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use:
-
IEC 61010-1, EN 61010-1
-
UL 61010-1, CSA 61010-1
Electromagnetic Compatibility
This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use; for radio equipment; and for telecommunication terminal equipment:
-
EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
-
EN 55011 (CISPR 11): Group 1, Class A emissions
-
EN 55022 (CISPR 22): Class A emissions
-
EN 55024 (CISPR 24): Immunity
-
AS/NZS CISPR 11: Group 1, Class A emissions
-
AS/NZS CISPR 22: Class A emissions
-
FCC 47 CFR Part 15B: Class A emissions
-
ICES-001: Class A emissions
Note:In the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial, light-industrial, and heavy-industrial locations. In Europe, Canada, Australia, and New Zealand (per CISPR 11) Class A equipment is intended for use only in heavy-industrial locations.
Note:Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis purposes.
CE Compliance
This product meets the essential requirements of applicable European Directives, as follows:
-
2014/35/EU; Low-Voltage Directive (safety)
-
2014/30/EU; Electromagnetic Compatibility Directive (EMC)